Model Number ES29162400
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 01/05/2018
Event Type  Injury  
Manufacturer Narrative
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information of conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event Description
According to the available information, tubing break at reinforced junction.
 
Manufacturer Narrative
This follow-up mdr is created to document the evaluation of the returned device: a titan touch pump and two cylinders were received for evaluation.Examination and testing of the returned components revealed a separation at the strain relief/exhaust tube junction of cylinder #2.Testing revealed this to be a site of leakage.The separation appears to be rough and irregular, indicating sufficient stress was exerted.Surface abrasion is noted on all strain relief and all tubes of the pump.No functional abnormalities are noted with the pump or cylinder #1.Based on quality's examination, quality concluded that while in-vivo both the exhaust tubes and inlet tube positioned themselves in such a way that caused them to overlap and abrade against one another.Quality further concluded that this positioning, in combination with device usage over time, could contribute to sufficient stress(s) to separate the exhaust tube of cylinder #2 at this site.A separation of this type would then allow the loss of fluid, making the device inoperable.Management routinely reviews events such as this and monitors complaint levels.Additionally, events of this type are captured in the product risk documentation.Based on this information no further corrective action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN TOUCH SCRO ZERO ANG 16CM
Type of Device
INFLATABLE PENILE PROSTHESIS
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebaek, 3050
DA  3050
Manufacturer (Section G)
COLOPLAST MANUFACTURING US, LLC
1601 west river road north
minneapolis MN 55411
Manufacturer Contact
stephanie perryman
1601 west river road north
minneapolis, MN 55411
6123024983
MDR Report Key7234289
MDR Text Key98781529
Report Number2125050-2018-00069
Device Sequence Number1
Product Code FHW
UDI-Device Identifier05708932539173
UDI-Public05708932539173
Combination Product (y/n)N
PMA/PMN Number
P000006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 02/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberES29162400
Device Catalogue NumberES29162400
Device Lot Number3951031
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2018
Was the Report Sent to FDA? No
Date Manufacturer Received01/05/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age67 YR
-
-