Catalog Number B1060-040 |
Device Problem
Torn Material (3024)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/12/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all relevant information.
|
|
Event Description
|
It was reported that the procedure was to treat a mildly tortuous and eccentric de novo distal superficial femoral artery.A 6.0 x 40 mm armada 35 balloon catheter was used; however, the balloon could not be inflated because it was noted to be torn.Therefore, the device was replaced with another unspecified armada balloon catheter to successfully complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
|
Manufacturer Narrative
|
Internal file number - (b)(4).Evaluation summary: visual inspection was performed on the returned device.The reported torn/ruptured balloon was confirmed.It was reported that the balloon was not prepped outside of the anatomy.It should be noted that the armada 35 instruction for use (ifu) instructs to prepare the balloon prior to introduction of the system into the anatomy.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other similar incidents.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
Event Description
|
Subsequent to the initial report, additional information was provided.Returned device analysis revealed that the balloon was inflated and there was a rupture observed.The device was inflated once at 20 atmospheres when it ruptured.No additional information was provided.
|
|