• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMPREHENSIVE REVERSE HUMERAL TRAY WITH LOCKING RING; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMPREHENSIVE REVERSE HUMERAL TRAY WITH LOCKING RING; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problems Bent (1059); Difficult To Position (1467)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/04/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4) there were multiple reports submitted for this event: mfr# 0001825034-2018-00291.Concomitant medical products: comprehensive reverse e1 humeral bearing pnep-115393 ln320400.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the patient underwent a reverse total shoulder arthroplasty.During the procedure, it was noticed that the locking ring would not hold the poly as it appeared to be bent.The deformation was observed as the surgeon was putting the poly into the humeral tray.The surgery was completed using another humeral tray.There was no surgical delay.No patient impact was reported.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual inspection revealed ringloc was bent within the tray.Visual inspection of the humeral tray indicates it is conforming to the print, except for the bend.There are no burrs or obstructions in the tray¿s ringloc groove.Dimensional inspections reveals measurements are within specifications.Device history record (dhr) was reviewed and no discrepancies were found relevant to the reported event.Root cause of the event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPREHENSIVE REVERSE HUMERAL TRAY WITH LOCKING RING
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7238402
MDR Text Key99092085
Report Number0001825034-2018-00290
Device Sequence Number1
Product Code KWS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK113069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number115370
Device Lot Number060730
Other Device ID Number(01) 0 0880304 54320 1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age77 YR
-
-