MAUDE Adverse Event Report: OBERDORF : SYNTHES PRODUKTIONS GMBH 4.9MM LOCKING BOLT 42MM ROD, FIXATION, INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OBERDORF : SYNTHES PRODUKTIONS GMBH 4.9MM LOCKING BOLT 42MM ROD, FIXATION, INTRAMEDULLARY Back to Search Results
Catalog Number 259.420
Device Problem Unintended Movement (3026)
Patient Problems Failure of Implant (1924); Unspecified Infection (1930)
Event Type  Injury  
Manufacturer Narrative
Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. Device history record review. Part number: 259. 420 synthes lot number: 1963481 release to warehouse date: 12. Aug. 2008 manufacturing site: mezzovico no ncrs were generated during production. Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition. A product development investigation was performed for the subject device. The present bolt was analysed for conformance to print specifications as well as the device history record was researched; no abnormal findings were identified. Manufacturing and inspection records indicated no problems with the lot in question. Based on these findings we conclude that the cause of failure is not due to any manufacturing non-conformance's. The relevant dimensions were measured and they have fulfilled the specifications. The bolt shows that the implant is in a used condition without heavy damage. The surface of this implant has slightly wear marks it is not possible to determine where it is coming from. We can only assume, that this can be traced during removal or a possible instability of mention loosen bolt ( according op-report; effect of local osteolysis). Measurement: document/drawing: smw_200235 feature/test/description: core diameter specification 4. 3 0/-0. 1 actual 4. 25 gage 3-01-19788 results "pass feature/test/description: outer diameter specification 4. 9 0/-0. 1 actual 4. 88 gage 3-01-19788 results "pass" feature/test/description: length specification 42 0/+0. 5 actual 42. 27 gage 3-01-19788 results "pass" based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.
 
Event Description
It was reported that a dislocation of the femoral neck blade occurred in the patient and there was a suspicion of a chronic low-grade infection. The bolts prove to be entirely loosened with no further significant grip on the bone. It is to be assumed that this effect is the result of local osteolysis.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Additional narrative: initially reported concomitant devices are no longer concomitant. Subject device has been received and is currently in the evaluation process. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported patient fell on the right side and suffered a dislocated intertrochanteric-subtrochanteric multiple fragment fracture of the femur, right side. Primary surgical treatment was not immediately possible due to patient¿s metformin intake. After easing the effects of the metformin, patient was implanted with a proximal femoral nail antirotation (pfna), a pfna blade, one (1) 38mm locking screw (bolt), one (1) 42 mm locking screw (bolt), and a pfna end cap on (b)(6) 2017. On unknown date, a dislocation of the femoral neck blade occurred in the patient, most notably through the re-opening of the sealed femoral neck blade. Accompanying this was a chronic persistent, opaque serous secretion over the entrance to the femoral neck blade. As a result, there was suspicion of a chronic low-grade infection. Patient was returned to surgery on (b)(6) 2017 where surgeon removed all hardware. During this procedure it was noted both screws (bolts) were loose with no significant grip on the bone. Surgeon assumed this to be the effect of local osteolysis. The wound was cleaned and the patient was closed. Unknown cultures were taken during this procedure.
 
Manufacturer Narrative
Patient¿s weight is unknown. Date of event is unknown. Additional device product codes: jdw, jds. The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes europe reports an event in (b)(6) as follows: it was reported that a patient fell on the right side and had a femoral fracture. Patient was treated with blade, nail and two locking bolts on (b)(6) 2017. Post-operatively, it was detected that the proximal femoral nail antirotation (pfna) blade has been unlocked and it was dislocated. Patient revised on (b)(6) 2017. During revision surgery it was also detected that both locking bolts were loose. Concomitant device reported: pfna 10 long r 130 l340 sst (part # 272. 280s, lot # 9377650, quantity 1), pfna end cap extens. 0 sst (part # 273. 155s, lot # l455811, quantity 1). This report is for one (1) 4. 9 mm locking bolt 42 mm. This is report 2 of 3 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name4.9MM LOCKING BOLT 42MM
Type of DeviceROD, FIXATION, INTRAMEDULLARY
Manufacturer (Section D)
OBERDORF : SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ 4436
MDR Report Key7241046
MDR Text Key253304260
Report Number8030965-2018-50848
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
PMA/PMN Number
K970097
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number259.420
Device Lot Number1963481
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2018
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/07/2018
Is This a Reprocessed and Reused Single-Use Device? No

Patient Treatment Data
Date Received: 02/05/2018 Patient Sequence Number: 1
-
-