The two cyf-va2 were returned to omsc for evaluation.
The evaluation found following failures.
[cyf-va2, s/n (b)(4), manufactured on september 13, 2015].
- there were kinks in the instrument channel and the insertion tube.
- there were debris and scratches inside of the instrument channel.
- there were scratches on the rubber of the bending section.
- the part of the coating over the insertion tube was peeling off.
[cyf-va2, s/n (b)(4), manufactured on november 11, 2010].
- there were kink in the instrument channel and the insertion tube, the universal cord and the instrument channel.
- there were scratches inside of the instrument channel.
- there were scratches on the rubber of the bending section and the adhesive of bending rubber.
- there was debris on the universal cord.
Omsc reviewed the manufacture history of the two cyf-va2 and confirmed no irregularity.
The exact cause could not be determined at present, if significant additional information is received, this report will be supplemented.
|
Olympus medical systems corp.
(omsc) was informed that 8 patients had a fever after a cystoscopy using subject device.
The 8 patients reportedly experienced cystoscopy between (b)(6) 2017 and (b)(6) 2018.
The facility provided unspecified medical treatment for the patients, but there was no information of the outcome.
The facility possesses two cyf-va2 (serial number: (b)(4)), but it is unknown which scope was used for each patient.
The user facility conducted microbial culture test for two cyf-va2 and the testing indicated no microbial growth for the two cyf-va2.
The user facility reported that the subject device had been reprocessed using endostream, a non-olympus automated endoscope reprocessor model, this report is 3 of 8 reports.
|