Model Number 508-32-103
Device Problem Contamination (1120)
Patient Problems Unspecified Infection (1930); No Code Available (3191)
Event Date 01/23/2018
Event Type  Injury  
Event Description
Revision surgery - due to the patient having an infection.The surgeon washed out and replaced the head and poly.
 
Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the revision detailed in this investigation occurred 2 years apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Type of Device
GLENOID, HEAD W/RETAINING SCREW, RSP, 32MM/-4MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7244053
MDR Text Key99122968
Report Number1644408-2018-00110
Device Sequence Number1
Product Code KWS
UDI-Device Identifier00888912024747
UDI-Public(01)00888912024747
Combination Product (y/n)N
PMA/PMN Number
K051075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 02/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/28/2021
Device Model Number508-32-103
Device Catalogue Number508-32-103
Device Lot Number864C2122
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
508-00-032, LOT 855C1946; 508-00-032, LOT 855C1946
Patient Outcome(s) Required Intervention;
Patient Age70 YR
-
-