DATEX-OHMEDA, INC. AESPIRE 7100 ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/05/2018
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the equipment and was not able to confirm the reported complaint. The bag/vent microswitch was replaced and the unit was returned to service. Patient information could not be obtained due to country privacy laws. The initial reporter is located outside the u. S. , and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported a failure of the unit to switch to mechanical ventilation when requested. There is no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAESPIRE 7100
Type of DeviceANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison WI 53718
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 53188
8479711808
MDR Report Key7244708
MDR Text Key99194005
Report Number2112667-2018-00250
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 02/06/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-