MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ANTHEM RF CRT-P IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ANTHEM RF CRT-P IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3210
Device Problem Operating System Version or Upgrade Problem (2997)
Patient Problems Muscle Stimulation (1412); Palpitations (2467)
Event Date 01/30/2018
Event Type  Injury  
Event Description
It was reported that during an attempted firmware upgrade, the pacemaker exhibited backup vvi. The patient experienced palpitations due to backup vvi pacing. After device download, normal device function was restored. The patient elected not to upgrade the firmware due to feeling muscle stimulation. The patient would continue to be followed normally.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameANTHEM RF CRT-P
Type of DeviceIMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7246224
MDR Text Key99177749
Report Number2017865-2018-02130
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation
Type of Report Initial
Report Date 02/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/30/2014
Device Model NumberPM3210
Device Lot Number3825222
Other Device ID Number05414734503785
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/30/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 02/06/2018 Patient Sequence Number: 1
-
-