• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASPEN SURGICAL PRODUCTS, CALEDONIA BARD-PARKER CARBON RIB-BACK BLADES SIZE 15, STERILE BARD-PARKER BLADE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASPEN SURGICAL PRODUCTS, CALEDONIA BARD-PARKER CARBON RIB-BACK BLADES SIZE 15, STERILE BARD-PARKER BLADE Back to Search Results
Model Number 371115-150
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem Foreign Body In Patient (2687)
Event Date 12/29/2017
Event Type  Injury  
Manufacturer Narrative
The distributor indicated in report that no sample was available for evaluation, only the manufacturing lot number was provided for review. The bard-parker blade size 15 broke during a knee arthroscopy procedure. The surgeon was able to retrieve the blade without incident. The patient tolerated well and follow up care was not required. Based on the evaluation conducted, lot 0146445 was packed per dhr and no non-conformance related to the broken blade were present. Packing process has established controls to mitigate broken blade condition, including a blade sensor that inspect 100% of packed pouches liner level prior to aluminum foil packaging. The following controls are in-place to mitigate ¿broken blade¿ condition at aspen surgical las piedras site: heat treatment in-process at the beginning and end of each lot are inspected for dimension integrity using a pin gauge, flatness gauge and perforation length gauge, ductility test, and hardness test. Heat treatment quality inspections at the beginning and end of each lot are inspected for dimension integrity using a pin gauge, flatness gauge and perforation length gauge, ductility test, and hardness test. Based on this information, no further action is required. Device not available.
 
Event Description
Aspen surgical received a report from the distributor indicating that a bard-parker blade broke off in the patient during a procedure. The incident occurred at the user facility. This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameBARD-PARKER CARBON RIB-BACK BLADES SIZE 15, STERILE
Type of DeviceBARD-PARKER BLADE
Manufacturer (Section D)
ASPEN SURGICAL PRODUCTS, CALEDONIA
6945 southbelt dr. s.e.
caledonia MI 49316
Manufacturer (Section G)
SAME AS ABOVE
Manufacturer Contact
jordan hackert
6945 southbelt dr. s.e.
caledonia, MI 49316
6165367508
MDR Report Key7248972
MDR Text Key99299590
Report Number1836161-2018-00008
Device Sequence Number1
Product Code GES
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation
Type of Report Initial
Report Date 01/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model Number371115-150
Device Lot Number0146445
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/08/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 02/07/2018 Patient Sequence Number: 1
-
-