Model Number UNITY PACS R11.1.2.5
Device Problems Computer Software Problem (1112); Loss of Data (2903)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/10/2018
Event Type  malfunction  
Manufacturer Narrative
Merge unity pacs is designed to display a message to the user when an exam report fails to upload to the image server.In this instance, there was no evidence that the physician received the message warning that the report did not upload.It was not possible to determine if the message displayed on the screen for the physician when the report was first read and saved.Unity investigation showed the database was updated but the report did not upload to the server.Audit trail showed the exam was read, approved and saved.The exam was re-read by the physician.Station logs had truncated, therefore were unavailable for review.No other review/investigation will be performed for cause.
 
Event Description
Merge unity pacs is a medical image and information management system that is used for viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems.On (b)(6) 2018, the customer reported an exam report was not saved and is missing.The unity tech investigation showed that the report did not upload successfully, and would have to be recreated.The audit trail does show a "report save" event.At the time of the investigation, station logs had truncated, and were unable to be reviewed.While images are available, a finalized report not being available for subsequent review by the patient's physician could potentially result in a delay in care that could lead to harm.However, the customer did not allege any harm, serious injury or death as a result of this issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE UNITY PACS
Type of Device
PICTURE AND ARCHIVING COMMUNICATIONS SYSTEM
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
meg mucha
900 walnut ridge drive
hartland, WI 53029
MDR Report Key7250545
MDR Text Key99323800
Report Number2183926-2018-00008
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041935
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNITY PACS R11.1.2.5
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-