Model Number 728246
Device Problems Signal Artifact/Noise (1036); Degraded (1153)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint pr# (b)(4).
 
Event Description
Engineering analysis concluded that this event has the potential to result in image misinterpretation due to an artifact from a degraded compensator within the collimator.Therefore, this issue has been determined to be a reportable event.
 
Manufacturer Narrative
The customer reported image artifacts on scanned images.There was no report of misrepresentation as a result of the artifacts.The philips field service engineer (fse) visually inspected the system, checked the raw data, and completed air calibrations to confirm the reported issue.The fse found a damaged compensator in the a-plane collimator as the cause of the image artifacts and replaced the a-plane collimator to resolve this issue.The fse confirmed there was no harm to a patient and no report of misrepresentation and/or mistreatment as a result of this reported issue.The system is in clinical use.This issue has been determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRILLIANCE 16 AIR
Type of Device
COMPUTED TOMOGRAPHY X-RAY
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7250674
MDR Text Key99315116
Report Number1525965-2018-00331
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
PMA/PMN Number
K012009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number728246
Device Catalogue NumberNCTA399
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/18/2016
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-