• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL,INC. 8 MM SMALL GRASPING RETRACTOR ENDOSCOPIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL,INC. 8 MM SMALL GRASPING RETRACTOR ENDOSCOPIC INSTRUMENT Back to Search Results
Model Number 470318-08
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/03/2018
Event Type  malfunction  
Manufacturer Narrative
Intuitive surgical, inc. (isi) received the instrument involved with this complaint and completed the device evaluation. Failure analysis confirmed the customer reported failure. The instrument was found to have a broken pitch cable at the distal clevis hub. The broken cable segment that contains the crimp was missing from clevis in the clevis. The clevis did not exhibit any damage or wear marks. Other cables at the wrist were not damaged. Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to this reported event. The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.
 
Event Description
It was reported that during a da vinci-assisted surgical procedure, a cable on the small grasping retractor instrument broke while grasping tissue. There was no report of fragment(s) falling inside the patient, patient harm, adverse outcome or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name8 MM SMALL GRASPING RETRACTOR
Type of DeviceENDOSCOPIC INSTRUMENT
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.,
950 kifer rd
sunnyvale CA
Manufacturer Contact
tabitha reed
950 kifer rd
sunnyvale, CA 
4085232420
MDR Report Key7251098
MDR Text Key99799675
Report Number2955842-2018-00083
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation
Type of Report Initial
Report Date 01/17/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number470318-08
Device Lot NumberN11170728 0023
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-