• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 QUICKSET ACE GRATER HEAD 56MM; HIP INSTRUMENTS : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 QUICKSET ACE GRATER HEAD 56MM; HIP INSTRUMENTS : REAMERS Back to Search Results
Catalog Number 244000556
Device Problem Material Discolored (1170)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/23/2018
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
After being processed by the sterilization dept, seems like there is rust on every acetabular reamer in the set.All acetabular reamer needs to be replace and investigated.Please send replacement to (b)(6).
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #
=
> pc-(b)(4).Investigation summary
=
> examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET ACE GRATER HEAD 56MM
Type of Device
HIP INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46582 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46582 0988
Manufacturer Contact
chad gibson
700 orthopaedic dr.
warsaw, IN 46582-0988
5743725905
MDR Report Key7252591
MDR Text Key99814242
Report Number1818910-2018-53070
Device Sequence Number1
Product Code HTO
UDI-Device Identifier10603295124030
UDI-Public10603295124030
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number244000556
Device Lot NumberSO2023156
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-