Model Number 72202682
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/30/2018
Event Type  malfunction  
Event Description
It was reported that while the patient was under traction and the doctor was operating the center strap of the active heel traction boot busted and the clasp broke.The doctor lost immediate traction.There were no instruments in the joint at the time.Procedure was delayed for more than two hours while getting a backup device to complete the procedure.No patient injury reported.
 
Manufacturer Narrative
An evaluation was performed by the supplier and could confirm the customer complaint for the velcro came off and the o-rings snapped around the ankle pad.A visual inspection was performed and showed the velcro was not returned.There is loctite present on the boot and applied within specification however without the velcro pad a complete evaluation can¿t be performed.The o-rings are there to help keep the pad in place while positioning the heel, but once the foot is in and secure they serve no function.The o-rings can break if pulled too hard.Smith & nephew will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AHTB ACTIVE HEEL TRACTION BOOT
Type of Device
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 
MDR Report Key7252900
MDR Text Key99607149
Report Number3003604053-2018-00023
Device Sequence Number1
Product Code HST
UDI-Device Identifier00885554024807
UDI-Public(01)00885554024807
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72202682
Device Catalogue Number72202682
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age20 YR
-
-