Model Number 42856
Device Problem Mechanics Altered (2984)
Patient Problem Hemorrhage, Subdural (1894)
Event Date 01/28/2018
Event Type  Injury  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report. If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported the patient developed a bilateral subdural hematoma and was admitted on (b)(6) 2018. The valve was originally implanted at a setting of 1. 5, however when checked during admission, the setting was at 0. 5. The valve was explanted the following day. After explant, the setting was checked and found to be at 2. 5.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTRATA II VALVE, SMALL
Type of DeviceSHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key7252905
MDR Text Key99566570
Report Number2021898-2018-00071
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Type of Report Initial
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number42856
Device Catalogue Number42856
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/29/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 02/08/2018 Patient Sequence Number: 1
-
-