Model Number 24970A
Device Problems Device Difficult to Program or Calibrate (1496); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/08/2017
Event Type  Malfunction  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

It was reported that during the implant procedure, the user attempted to change the pacing mode on the mobile programmer application as a precaution and received an analyzer electrical reset warning. The application was closed, re-launched, the mobile device re-paired with the programmer base, and the analyzer restarted three times before this mode switch was able to be completed, with several unsuccessful attempts made between each restart. It was noted that the programmer base was sitting on a metal table and was in the vicinity of electro-cautery. The programmer base and application remain in use. No patient complications have been reported as a result of this event.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARELINK SMARTSYNC
Type of DeviceANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7254837
MDR Text Key99834973
Report Number3004593495-2018-00043
Device Sequence Number1
Product Code DTC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK163008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation
Type of Report Initial
Report Date 02/08/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received02/08/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number24970A
Device Catalogue Number24970A
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/08/2017
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured03/23/2017
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-