MEDTRONIC EUROPE SARL REVEAL LINQ DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problems Invalid Sensing (2293); Device Sensing Problem (2917)
Patient Problems Seizures (2063); No Known Impact Or Consequence To Patient (2692)
Event Date 08/30/2017
Event Type  Malfunction  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

It was reported that the patient passed out and "had seizure like activity" during a pause episode. It was noted that the patient manually marked this pause symptom, but multiple pause episodes were not detected due to the implantable cardiac monitor (icm)'s algorithm and that the device was sensing p-waves as r-waves. The device has been explanted and the patient had a pacemaker implanted. No patient complications have been reported as a result of this event.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameREVEAL LINQ
Type of DeviceDETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7256731
MDR Text Key99899429
Report Number9614453-2018-00425
Device Sequence Number1
Product Code DSI
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation HEALTH PROFESSIONAL
Type of Report Initial
Report Date 02/08/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received02/08/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date03/28/2018
Device MODEL NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/12/2017
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured03/31/2017
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-