SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 36MM X 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Catalog Number 71335752
Device Problem Loose or Intermittent Connection (1371)
Patient Problem Joint Dislocation (2374)
Event Date 01/19/2018
Event Type  Injury  
Event Description
Patient positioned and hip dislocated.Liner and screws removed cup was loose.Removed cup reamed acetabulum and implanted a 52mm multihole cup and three screws.A constrained liner inserted and checked to be locked in the cup.Hip reduced using a 22mm stryker head.Hip stable.
 
Manufacturer Narrative
The associated complaint device was not returned for evaluation the associated r3 3hole acetabular shell, r3 acetabular liner, head screws and hole cover were not returned for evaluation.A review of complaint history on the listed parts revealed no additional complaints for this issue with the same batch number.Our investigation included a review of the manufacturing records which did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A clinical evaluation noted that based on the available information, the root cause of the ¿cup loosening¿ cannot be concluded.It is unknown if the post-primary surgery fall contributed to the ¿loosening¿ but that explanation cannot be excluded.Without the return of the actual product involved, our investigation is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG XLPE ACET LNR 36MM X 52MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
Manufacturer Contact
markus poettker
1450 brooks road
memphis, TN 38116
MDR Report Key7262494
MDR Text Key99762842
Report Number1020279-2018-00151
Device Sequence Number1
Product Code MBL
UDI-Device Identifier03596010598332
UDI-Public03596010598332
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K113848
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Catalogue Number71335752
Device Lot Number17GM12373
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/22/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight150
-
-