• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.710
Device Problems Dull, Blunt (2407); Output below Specifications (3004)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/25/2018
Event Type  malfunction  
Manufacturer Narrative
The reporter¿s phone number and complete address were not provided.The manufacturing site name and location were unknown.The serial number was unknown.Therefore, device manufacture date is unknown.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, the surgeon discovered that the movement of the battery oscillator device was dull.There was a five minute delay to the surgical procedure.A spare device was used to complete the surgery.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Serial number: the device serial number was unknown in the initial report.Upon receipt of the device, the serial number was identified as (b)(4).The udi, date of manufacturer and manufacture site name and address have been updated accordingly.Device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.It was further determined that the trigger was blocked, jammed and was moving heavily.It was further determined that the locking ring of the trigger had become loose, preventing the trigger from being pressed to the end-stop position.As a result, the maximum speed could not be reached (lack of performance).It was further determined that the device failed pretest for trigger test, check the triggers and electronic control unit (ecu) function and check oscillation frequency 10800 - 14299 oscillation/minute.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR II FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7268119
MDR Text Key100266614
Report Number8030965-2018-51081
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819491847
UDI-Public(01)07611819491847
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.710
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2018
Date Manufacturer Received03/14/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-