It was reported that during a laparoscopic myomectomy, the tip of the outer sleeve was bend after a myoma borer was inserted.Then, the tip of the outer sleeve was broken off when the device was removed from the patient.The device was used on the uterus.No pieces were left inside the patient.Another device was used to complete the case.There were no adverse consequences to the patient.
|
(b)(4).Batch # p9430u, p93y7x.Device analysis: the analysis results found that the b5st instrument was received with the sleeve melted.In addition, the tyvek was returned along with the instrument.This type of damage is consistent with the one produced by an energized device.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|