Catalog Number B5ST
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/22/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event Description
It was reported that during a laparoscopic myomectomy, the tip of the outer sleeve was bend after a myoma borer was inserted.Then, the tip of the outer sleeve was broken off when the device was removed from the patient.The device was used on the uterus.No pieces were left inside the patient.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Batch # p9430u, p93y7x.Device analysis: the analysis results found that the b5st instrument was received with the sleeve melted.In addition, the tyvek was returned along with the instrument.This type of damage is consistent with the one produced by an energized device.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7270437
MDR Text Key100247492
Report Number3005075853-2018-07992
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier20705036001239
UDI-Public20705036001239
Combination Product (y/n)N
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2022
Device Catalogue NumberB5ST
Device Lot NumberP4T71U
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2018
Date Manufacturer Received02/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-