ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); Retraction Problem (1536)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/01/2018
Event Type  malfunction  
Event Description
It was reported that during implant the physician had a difficulty removing the stylet.It was observed that the physician used moderate force to remove the stylet, once removed it was noted the pin on the lead was detached.The lv lead was removed and new lead placed.The patient was stable.
 
Manufacturer Narrative
The reported field event of the inability to remove the stylet was confirmed.Visual inspection of the stylet noted the coating was bunched with the inner coil at the distal end of the connector pin.The coating anomaly would have led to difficulty removing the stylet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7270547
MDR Text Key100073590
Report Number2938836-2018-01251
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2020
Device Model Number1458Q/86
Device Lot NumberA000048485
Other Device ID Number05414734503198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age87 YR
-
-