STERIS MEXICO, S. DE R.L. DE C.V. 20" CENTURY STERILIZER Back to Search Results
Device Problem Fluid Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/16/2018
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite to inspect the sterilizer. The technician inspected the sterilizer and found that a heating element on the sterilizer's integral steam generator was fractured subsequently allowing water to leak out. The technician repaired the sterilizer's steam generator, ran a test cycle and confirmed the unit to be operating properly. The sterilizer was returned to service and no additional issues have been reported.
 
Event Description
The user facility reported that water was leaking from their sterilizer. No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name20" CENTURY STERILIZER
Type of DeviceSTERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX 67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX 67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key7272020
MDR Text Key100284540
Report Number3005899764-2018-00011
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/14/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received01/16/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-