Model Number 1MTEC30 |
Device Problem
Scratched Material (3020)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
If implanted, give date: not applicable, as the product is not implantable.If explanted, give date: not applicable, as the product is not implantable.All pertinent information available to abbott medical optics has been submitted.
|
|
Event Description
|
It was reported that the doctor believes the (b)(4) cartridges associated with lot number cc13687 are causing the intraocular lenses to scratch on insertion with no patient contact.Reportedly, the problem with these cartridges is not because of use error.The scratches are at the far periphery underside of the iol on insertion and not in the area with contact with the plunger and not in the area with the loading forceps.No additional information was provided.
|
|
Manufacturer Narrative
|
Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation, there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
|
Search Alerts/Recalls
|