BECTON, DICKINSON & CO., (BD) BD VACUTAINER® WINGED SAFETY PUSH BUTTON BLOOD COLLECTION SET
|
Back to Search Results |
|
Catalog Number 367342 |
Device Problems
Retraction Problem (1536); Structural Problem (2506)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/23/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation and/or device history review, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the needle of a bd vacutainer winged safety push button blood collection set failed to retract, then ¿fell apart¿.There was no report of exposure, injury or medical intervention.
|
|
Manufacturer Narrative
|
Investigation summary: bd received samples from the customer facility for investigation.The samples were tested and the customer's indicated failure mode for non-retraction with the incident lot was not observed as all product specifications were met.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for non-retraction with the incident lot was not observed as all samples met the required specifications.Root cause description: based on the investigation, a root cause could not be determined.The product was found to be in conformance and meet release specifications.
|
|