(b)(4).It has been communicated that the device and/or lot information is not available for evaluation.Without the device and/or lot information it is not possible for codman to conduct a proper investigation.If at some point the device and/or lot information does become available, this complaint will be re-opened, evaluated and a follow up report will be filed.Trends will be monitored for this and similar complaints.At the present time this complaint is considered closed.
|
Updated udi : (b)(4).The device was returned for evaluation.The valve was visually inspected; a small cut/tear was noted in the silicone housing over the valve casing and needle holes in the needle chamber, no biological debris was noted.The position of the cam when valve was received was 70mmh2o.The valve was tested for programming.The valve failed the test, the cam mechanism did not move during the programming process.The valve was flushed; no occlusion was noted.The valve was leak tested; only leaked from the needle holes in the needle chamber, did not leak from the small cut/tear in the silicone housing.The catheters were irrigated, no occlusions noted.The valve was reflux tested and passed.The siphon guard was tested and passed.The siphon guard was removed and the valve was pressure tested; the valve passed.The cam mechanism was moved with a needle, no resistance was noted.The valve was program tested and failed; the cam did not move.The cam magnets were controlled and failed testing.The magnets polarity was controlled and failed; all magnets were on (+).Review of the device history records for the valve found the device conformed to specification when released to stock.The root cause for the cut/tear in the silicone housing is due to a sharp or pointed object coming into contact with the silicone, as noted in the ifu silicone has a low tear and cut resistance.The root cause for the programming problem is due to the cam magnets.The abnormal polarization of the valve was likely caused by exposure to too strong magnetic field.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|