MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION COMPACT
Device Problems Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 01/25/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern. The interface box for the navigation system was returned to the manufacturer for evaluation. Testing found that the box was unable to establish communication with a known operational navigation system. The hardware investigation found that the reported event was related to a hardware issue. This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Event Description
A medtronic representative reported that, while outside of a procedure, the fault light on the display panel was illuminated. It was reported that the communication cables were swapped, the navigation system restarted and the emitters swapped without resolution. There was no patient present when this issue was identified. No additional information was provided.
 
Manufacturer Narrative
Additional information: a medtronic representative reported that replacement of the interface box of the navigation system restored functionality. Correction: unique device identification (udi) and device manufacture date updated to proper value.
 
Manufacturer Narrative
Additional information: a medtronic representative went to the site to test the equipment. The hardware, software, and instruments passed the system checkout. The system was found to be fully functional.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMEDTRONIC FUSION COMPACT NAVIGATION SYSTEM
Type of DeviceEAR NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
heather davis
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7280231
MDR Text Key100539352
Report Number1723170-2018-00724
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 05/24/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberFUSION COMPACT
Device Catalogue Number9735602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-