PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. BRILLIANCE AIR 40/64/UCT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
|
Back to Search Results |
|
Device Problem
Signal Artifact/Noise (1036)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.(b)(4).
|
|
Event Description
|
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported ring artifacts displayed in ct head images and a rescan was performed.A philips field service (fse) confirmed the compensator was cracked and was causing the intermittent ring artifacts.Engineering analysis has confirmed that this event has been determined to be a reportable issue due to the potential for image misinterpretation, because of an artifact.Therefore, this issue has been determined to be a reportable event.This event is currently under investigation.
|
|
Manufacturer Narrative
|
The customer reported having ring artifacts on patient images.The philips field service engineer (fse) evaluated the system and confirmed there was a crack in the compensator.The fse replaced the a-plane collimator to resolve the issue.The philips fse confirmed there was no harm to a patient, operator, or bystander.The system is in clinical use.This issue was determined not to be a reportable event.
|
|