PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. BRILLIANCE AIR 40/64/UCT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Device Problem Signal Artifact/Noise (1036)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.(b)(4).
 
Event Description
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported ring artifacts displayed in ct head images and a rescan was performed.A philips field service (fse) confirmed the compensator was cracked and was causing the intermittent ring artifacts.Engineering analysis has confirmed that this event has been determined to be a reportable issue due to the potential for image misinterpretation, because of an artifact.Therefore, this issue has been determined to be a reportable event.This event is currently under investigation.
 
Manufacturer Narrative
The customer reported having ring artifacts on patient images.The philips field service engineer (fse) evaluated the system and confirmed there was a crack in the compensator.The fse replaced the a-plane collimator to resolve the issue.The philips fse confirmed there was no harm to a patient, operator, or bystander.The system is in clinical use.This issue was determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRILLIANCE AIR 40/64/UCT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7280499
MDR Text Key100549822
Report Number1525965-2018-00353
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/22/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-