Model Number ORT100
Device Problems Mechanical Problem (1384); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/31/2018
Event Type  malfunction  
Manufacturer Narrative
The service engineer observed that the roll cylinder in the table was leaking.He removed and replaced the cylinder and tested the table to verify proper function.Internal capa issued.
 
Event Description
Prior to the procedure, the neuro lead noticed that the table top was tilted supine patient right.He informed the surgeon and the staff.They proceeded to bring the patient into the room and set up the room.The patient was in the supine position and non-pinned.Within a short time, the operating room (or) staff observed that the tilt of the table top was more substantial.They elected to move the patient to a different or and complete the procedure there.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMRIS ORT100 OPERATING ROOM TABLE
Type of Device
OPERATING ROOM TABLE
Manufacturer (Section D)
IMRIS - DEERFIELD IMAGING, INC.
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer (Section G)
IMRIS - DEERFIELD IMAGING, INC.
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer Contact
todd sperling
5101 shady oak road
minnetonka, MN 55343-4100
7632036344
MDR Report Key7281229
MDR Text Key100567310
Report Number3010326005-2018-00023
Device Sequence Number1
Product Code KXJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Remedial Action Replace
Type of Report Initial
Report Date 02/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Model NumberORT100
Device Catalogue Number110470-000
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/31/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/15/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-