DATEX-OHMEDA, INC. AESPIRE 7100 ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 01/23/2018
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint. The subminiature switch was replaced, and the unit was returned to service. No report of patient involvement. The initial reporter is located outside the u. S. , and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported an intermittent failure of the unit to switch to mechanical ventilation when requested, during preuse checkout. There is no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAESPIRE 7100
Type of DeviceANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7284330
MDR Text Key100579253
Report Number2112667-2018-00374
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 02/20/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received01/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-