Model Number 8637-20
Device Problem Premature Elective Replacement Indicator (1483)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/30/2016
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient receiving unknown drug via an implanted pump. The indication for use was intractable spasticity and cerebral palsy. It was reported the patient's pump was interrogated on (b)(6) 2016 and it noted 16 months until eri. The pump was then interrogated on (b)(6) 2016 and it noted eri occurred on (b)(6) 2016. The pump needed to be replaced before (b)(6) 2017. The patient's pump was explanted on (b)(6) 2017.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additionalinformation was received from a healthcare provider on 2018-mar-09. It was reported that the patient did not experience any symptoms related to the premature eri, and it was specified that there were no withdrawal symptoms. Regarding the cause of the premature eri, it was noted that eri went from 15 months to 90 days since the patient's last refill on (b)(6) 2016. It was noted that the issue was not resolved until replacement on (b)(6) 2017 (note: this conflicts with the previous report that the pump was explanted on (b)(6) 2017). No further complications were reported or anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7286598
MDR Text Key100600807
Report Number3004209178-2018-03711
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date08/28/2012
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/09/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured03/10/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 02/21/2018 Patient Sequence Number: 1
-
-