Model Number 8637-20 |
Device Problem
Premature Elective Replacement Indicator (1483)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/30/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a consumer regarding a patient receiving unknown drug via an implanted pump.
The indication for use was intractable spasticity and cerebral palsy.
It was reported the patient's pump was interrogated on (b)(6) 2016 and it noted 16 months until eri.
The pump was then interrogated on (b)(6) 2016 and it noted eri occurred on (b)(6) 2016.
The pump needed to be replaced before (b)(6) 2017.
The patient's pump was explanted on (b)(6) 2017.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additionalinformation was received from a healthcare provider on 2018-mar-09.
It was reported that the patient did not experience any symptoms related to the premature eri, and it was specified that there were no withdrawal symptoms.
Regarding the cause of the premature eri, it was noted that eri went from 15 months to 90 days since the patient's last refill on (b)(6) 2016.
It was noted that the issue was not resolved until replacement on (b)(6) 2017 (note: this conflicts with the previous report that the pump was explanted on (b)(6) 2017).
No further complications were reported or anticipated.
|
|