• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-7008A
Device Problems False Reading From Device Non-Compliance (1228); Device Displays Incorrect Message (2591)
Patient Problem Hyperglycemia (1905)
Event Date 02/12/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 188, 138 and 289 mg/dl and the sensor glucose was 66 and 44 mg/dl.The customer¿s current blood glucose was 198 mg/dl.Insulin delivery was suspended due to sensor values.Sensor value that triggered the suspend event was 40 mg/dl.Suspend on low limit in sensor settings was 60 mg/dl.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The sensor will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008A
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7289758
MDR Text Key100708488
Report Number3004209178-2018-52547
Device Sequence Number1
Product Code OZO
UDI-Device Identifier20643169646432
UDI-Public(01)20643169646432(017)20180502
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/02/2018
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Device Lot NumberHG23Z6B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/12/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age52 YR
Patient Weight155
-
-