Model Number MMT-751NAS |
Device Problems
No Display/Image (1183); Fluid/Blood Leak (1250)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/08/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer was reported via phone call that, the insulin pump had partial display screen.The blood glucose was 142 mg/dl at the time of the incident.Customer fell while shoveling and then notice condensation on the screen it was working and then it was blinking and then stop working.Customer reported fluid in the battery compartment and home screen appeared on the display.Troubleshooting determined that, the insulin pump should be replaced.The customer advised to discontinue use of the pump and revert to a back-up plan.The device will be returned for analysis.
|
|
Manufacturer Narrative
|
Pump passed the operating currents measurement, self test and displacement test.No missing segments/partial display noted.Pump was monitored for several days and no blank display anomaly, display anomaly or flashing on and off anomaly noted.No damage found on the lcd isolation tape noted.No moisture damage found inside the pump.
|
|