MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 BASIC SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 BASIC SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problems Device Stops Intermittently (1599); Communication or Transmission Problem (2896)
Patient Problem No Patient Involvement (2645)
Event Date 05/10/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.It was reported that the camera communication cable was replaced.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The cable was returned to the manufacturer for evaluation.Testing found that the cable had an intermittent open on one pin within the connector.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Event Description
A medtronic representative reported that, while outside of a procedure, the camera of the navigation system was not communicating with the main system.There was no patient present when this issue was identified.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 BASIC SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7299330
MDR Text Key101167187
Report Number1723170-2018-00876
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722194
UDI-Public00643169722194
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 02/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735665
Is the Reporter a Health Professional? No
Date Manufacturer Received02/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/09/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-