• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 382400
Device Problem Application Program Problem: Dose Calculation Error (1189)
Patient Problem No Patient Involvement (2645)
Event Date 01/18/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that an issue occurred with an enteral feeding pump.Upon triage on (b)(6) 2018 the service tech found that the device displayed that it fed above tolerance during a volumetric accuracy testing.The device was set to run at 150ml/hr and would do so for 30 minutes to complete the volumetric accuracy test.The device delivered 72.72ml of fluid during the test, measured by mass and converted into ml.The device displayed that it fed 88ml of fluid.The device delivers a proper amount but displays that it fed over the proper tolerance.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed and the customer states ¿the device has a low reading.¿ the unit was triaged and the reported issue could not be confirmed at this time.The unit delivered accurate amounts of food.However, the unit displayed an incorrect amount of food volume was fed.Therefore, the root cause was isolated to a software issue.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
Manufacturer (Section G)
COVIDIEN
building 10- no 789 puxing roa
shanghai 20111 4
CN   201114
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7300759
MDR Text Key101158002
Report Number3006451981-2018-00135
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521006614
UDI-Public10884521006614
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/26/2022
Device Model Number382400
Device Catalogue Number382400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received02/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-