Model Number 728306 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.(b)(4).
|
|
Event Description
|
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported that the operator¿s voice cannot be heard in the scan room.Philips service also reported that the operator could not hear the patient via the intercom.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.This event is currently under investigation.
|
|
Manufacturer Narrative
|
The customer reported that the operator¿s voice cannot be heard in the scan room.Philips service also reported that the operator could not hear the patient via the intercom.The customer confirmed there was no patient impact and no harm as a result of this event.The philips field service engineer (fse) went to the customer site the same day to evaluate the ct system.The fse confirmed the customer¿s allegation that no voice could be heard in the scan room and also determined, through troubleshooting, that voices from the scan room could not be heard in the control room.The fse opened the ct gantry and checked the audio cable connections and re-seated them, and checked for sound; still, no voices could be heard.The fse determined there was an ict gantry audio fault.The fse replaced the front and rear breathing light module and the audio board to correct and resolve the issue.The system is operational and in clinical use.
|
|
Search Alerts/Recalls
|
|
|