Model Number 728306
Device Problem Defective Component (2292)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.(b)(4).
 
Event Description
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported that the operator¿s voice cannot be heard in the scan room.Philips service also reported that the operator could not hear the patient via the intercom.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.This event is currently under investigation.
 
Manufacturer Narrative
The customer reported that the operator¿s voice cannot be heard in the scan room.Philips service also reported that the operator could not hear the patient via the intercom.The customer confirmed there was no patient impact and no harm as a result of this event.The philips field service engineer (fse) went to the customer site the same day to evaluate the ct system.The fse confirmed the customer¿s allegation that no voice could be heard in the scan room and also determined, through troubleshooting, that voices from the scan room could not be heard in the control room.The fse opened the ct gantry and checked the audio cable connections and re-seated them, and checked for sound; still, no voices could be heard.The fse determined there was an ict gantry audio fault.The fse replaced the front and rear breathing light module and the audio board to correct and resolve the issue.The system is operational and in clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
Manufacturer Contact
derrick massuri
595 miner rd
cleveland, OH 44143
4404833000
MDR Report Key7303439
MDR Text Key101278037
Report Number1525965-2018-00363
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K060937
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number728306
Device Catalogue NumberNCTC780
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/31/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-