TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
|
Back to Search Results |
|
Model Number 816571 |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4) - no consequences or impact to patient.(b)(4).The field service representative (fsr) could not duplicate the reported complaint.He replaced the display and the pump cable.The unit operated to the manufacturer's specifications.The suspect parts were returned to the manufacturer for evaluation.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the display on the pump was intermittently flickering.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was not confirmed.During laboratory analysis, the product surveillance technician (pst) observed the display assembly and cable to function normally throughout evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|