• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC0303
Device Problems Break (1069); Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/05/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report the suspected cable break.It was reported that this was a mitraclip procedure to treat grade 4 functional mitral regurgitation (mr).The tip of the dilator and the steerable guide catheter (sgc) had barely been inserted into the femoral vein when a pop was heard.The tip no longer deflected with use of the plus/minus knob.A cable break was suspected.The sgc was replaced with a new one.The procedure continued with the successful implantation of one mitraclip, reducing mr to 1-2.There were no adverse patient effects and no clinically significant delay in the procedure.There was no additional information provided.
 
Manufacturer Narrative
(b)(4).All available information was investigated and the reported cable break and resultant tip deflection issue ((mechanical issue of steerable guide catheter (sgc) unable to curve guide)) were confirmed via returned device analysis.The reported noise could not be replicated in the testing environment as the event was a symptom of the cable break.A review of the lot history record revealed no manufacturing nonconformities.Additionally, a review of the complaint history identified no similar incidents reported from this lot.The investigation determined that the reported noise and mechanical issue of sgc unable to curve guide were a result of the cable break; however, a definitive cause for the cable break could not be definitively determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
MENLO PARK, CA REG# 3005070406
abbott vascular
26531 ynez rd.
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key7304569
MDR Text Key101356103
Report Number2024168-2018-01456
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/11/2018
Device Catalogue NumberSGC0303
Device Lot Number70710U137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient Weight90
-
-