Catalog Number SGC0303 |
Device Problems
Break (1069); Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is filed to report the suspected cable break.It was reported that this was a mitraclip procedure to treat grade 4 functional mitral regurgitation (mr).The tip of the dilator and the steerable guide catheter (sgc) had barely been inserted into the femoral vein when a pop was heard.The tip no longer deflected with use of the plus/minus knob.A cable break was suspected.The sgc was replaced with a new one.The procedure continued with the successful implantation of one mitraclip, reducing mr to 1-2.There were no adverse patient effects and no clinically significant delay in the procedure.There was no additional information provided.
|
|
Manufacturer Narrative
|
(b)(4).All available information was investigated and the reported cable break and resultant tip deflection issue ((mechanical issue of steerable guide catheter (sgc) unable to curve guide)) were confirmed via returned device analysis.The reported noise could not be replicated in the testing environment as the event was a symptom of the cable break.A review of the lot history record revealed no manufacturing nonconformities.Additionally, a review of the complaint history identified no similar incidents reported from this lot.The investigation determined that the reported noise and mechanical issue of sgc unable to curve guide were a result of the cable break; however, a definitive cause for the cable break could not be definitively determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
|
Search Alerts/Recalls
|