• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS NANOKNIFE SYSTEM; LOW ENERGY DIRECT CURRENT ABLATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS NANOKNIFE SYSTEM; LOW ENERGY DIRECT CURRENT ABLATION SYSTEM Back to Search Results
Model Number 20300101
Device Problem Device Stops Intermittently (1599)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 02/07/2018
Event Type  Injury  
Manufacturer Narrative
It was reported that the nanoknife system ((b)(4) ) involved in the incident is available to be returned to the manufacturer for evaluation.To date, the generator has yet to be returned.Attempts are being made to obtain the unit.The results of the unit evaluation will be sent via a follow up medwatch.(b)(4).
 
Event Description
As reported to angiodynamics on (b)(6) 2018, the treating physician reported that the irreversible electroporation (ire) generator shut down approximately 5-6 times during an ire procedure.As a result, the treatment is prolonged approximately 90 minutes longer than as normally performed.During this time, the patient is under anesthesia.It was reported that the patient suffered no adverse effects due to the event.The customer has requested the facility's unit be evaluated by the manufacturer.The unit has yet to be returned to the manufacturer for evaluation.
 
Manufacturer Narrative
The reported nanoknife unit was returned to angiodynamics' facility in (b)(4).Returned with the unit were the foot pedal and power cord.A visual examination of the unit noted it was in good physical condition.During functional testing, several rfid card errors were noted in the unit's logs.This indicated the need to replace the rfid board.The rfid board was replaced, however, when testing the unit, errors still appear in log files and ire shut down.Hardware services replaced the ac-dc board, the switching board, the charging board, and the "gome" board, however errors still appeared.It was determined that the issue was most likely due to a faulty motherboard or from the power supply since these components had not been replaced.As these components are not available to angiodynamics hardware services for replacement, angiodynamics is unable to continue to evaluate the unit.It was determined to return the unit to the manufacturer, igea, for inspection.Evaluation of the unit by igea determined the cmos battery was exhausted.The battery was replaced.Igea reset the bios and restored the bios with correct parameters.During the servicing of the unit, igea also noted the rfid board's sd card was not correctly inserted in it's slot.Igea correctly inserted the sd card.Igea determined the root cause for the unit failure was an exhausted battery and incorrect rfid board sd card.The battery was replaced and the rfid sd card was correctly inserted.The unit was returned to angiodynamics where it was tested.The unit passed all testing.Unit meets all acceptance criteria.A review of the device history records was performed for the (b)(4).The review confirms that the unit met all material, assembly, and performance specifications at the time of manufacture.The user manual, which is supplied to the user with this unit, contains the following statements: "output current restriction: when the generator senses that the current between the electrodes exceeds the operating parameters, the pulses are aborted.This safety feature protects against applying output energy that exceeds maximum current settings.Do not use the generator if a malfunction is suspected.Contact the manufacturer or the local authorized supplier.5.1.1 procedural guidelines: the nanoknife system requires muscle blockade and general anesthesia.0/4 to 1/4 twitches on the train of four test.Asa guidelines require a defibrillator be readily available for general anesthesia (defibrillator pads recommended)." a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NANOKNIFE SYSTEM
Type of Device
LOW ENERGY DIRECT CURRENT ABLATION SYSTEM
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
MDR Report Key7305633
MDR Text Key101240564
Report Number1319211-2018-00018
Device Sequence Number1
Product Code OAB
Combination Product (y/n)N
PMA/PMN Number
K102329
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model Number20300101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2018
Was the Report Sent to FDA? No
Date Manufacturer Received02/07/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-