• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAL
Device Problems Air Leak (1008); Crack (1135); Fluid/Blood Leak (1250); Incorrect Or Inadequate Test Results (2456)
Patient Problems Hyperglycemia (1905); Hypoglycemia (1912)
Event Date 02/08/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer was reported via phone call that, the customer had high and low blood glucose issues.Customer declined troubleshooting for the high blood glucose.Customer stated that, the location of the leak was in tubing.The blood glucose was over 600 mg/dl.Customer stated that, she noticed air bubbles is the tubing.Approximate size of air bubbles is unknown.Air bubbles were noticed by customer during therapy.Customer treated high blood glucose with manual injection syringe.The blood glucose was 48 mg/dl at the time of the incident and current blood glucose was unknown and today the customer noticed the tubing was cracked again blood glucose was 150 mg/dl.Customer reported the following symptoms related to their high blood glucose were nausea, vomiting.Based on customer report proceeding in troubleshooting per customer alleging possible pump under delivery.The device will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAL
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7307719
MDR Text Key101316939
Report Number3004209178-2018-54052
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169503694
UDI-Public(01)00643169503694
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-751NAL
Device Catalogue NumberMMT-751NAL
Device Lot NumberA4751NALJ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/14/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age47 YR
Patient Weight160
-
-