Catalog Number IGTCFS-65-JUG
Device Problems Unintended Movement (3026); Appropriate Term/Code Not Available (3191)
Patient Problems Internal Organ Perforation (1987); Perforation of Vessels (2135); Blood Loss (2597)
Event Type  Injury  
Manufacturer Narrative
William cook (b)(4) initially reported event under manufacturer report #3002808486-2017-02536.New information was received identifying that the product was a cook inc.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
Pt allegedly received an implant on (b)(6) 2010 via the right internal jugular vein due to pulmonary embolism, vein thrombosis.Pt is alleging tilt, vena cava perforation, perforation in the mesentery, bleeding in the lungs, shortness of breath and fatigue.The patient further alleges anxiety and stress.
 
Manufacturer Narrative
Additional information investigation - it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'gunther - tilt, vc & organ perf, bleeding in lungs, shortness of breath, fatigue, anxiety, stress - updated sfc.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Unknown if the reported bleeding in lungs, shortness of breath, fatigue, anxiety, stress is directly related to the filter and unable to identify corresponding failure mode(s) at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.H3 other text : blank fields on this form indicate the information is unknown or unavailable, or unchanged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP JUGULAR VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7308237
MDR Text Key101341989
Report Number1820334-2018-00618
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K072240
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-JUG
Was Device Available for Evaluation? No
Date Manufacturer Received04/19/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Weight118
-
-