DEPUY SYNTHES PRODUCTS LLC OSCILLATING SAW ATCH, LARGE, WITH KEY; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
Catalog Number 532.026 |
Device Problem
Output below Specifications (3004)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/06/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
This is report 1 of 2 for the same event.It was reported that during a veterinary tibial plateau leveling osteotomy surgery, it was observed that the small battery drive device and the oscillating saw with key had low power while being used together.It was not reported if there were any delays in the procedure.It was reported that they were able to successfully complete the surgery with the same devices.There was no human patient involvement as this was a veterinary procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.During repair, it was determined that the bearing on the device was seized, jammed, and moving heavily.It was further determined that the needle bearings were blocked.It was also determined that the device failed pre-test for check oscillation frequency.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to normal wear.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|