Catalog Number 106015
Device Problem Partial Blockage (1065)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/06/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4). Approximate age of device ¿ 2 years and 9 months. No further information was provided. A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Event Description
The patient was implanted with a left ventricular assist device (lvad) on (b)(6) 2015. It was reported that the patient presented with persistent low flow alarms. The alarm continued to sound while the patient was at the center. Labs were pending, results were not provided. The outflow graft (ofg) was to be reevaluated based on the last echocardiogram/ct scan findings. The cause of the low flow alarms was reported as of occlusion. The patient had significant stenosis found within previous ct angiography (cta) from an unspecified date which showed one focal area. The repeat cta during the this admission found multiple areas. The patient underwent stenting of the ofg. On (b)(6) 2018 it was reported that the patient was home, and the event had been resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEARTMATE II LVAS
Type of DeviceLEFT VENTRICULAR ASSIST DEVICE
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6101 stoneridge dr.
pleasanton, CA 94588
7818528390
MDR Report Key7312774
MDR Text Key101479374
Report Number2916596-2018-00822
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date10/31/2017
Device Catalogue Number106015
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/06/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/13/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

Patient Treatment Data
Date Received: 03/05/2018 Patient Sequence Number: 1
-
-