(b)(4) batch # p9454r, p94197, p9404h.Device analysis: the analysis results found that the tt012 device was received with the cannula broken; the broken piece of the cannula was returned inside a plastic bag.In addition, the tyvek was returned along with the instrument.One possible cause for the damage found on the cannula, may be excessive external load placed on the device.The lot history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the packaging process.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.Performed by hugo serrano.
|