• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASPEN SURGICAL PRODUCTS, CALEDONIA BARD-PARKER CARBON RIB-BACK BLADES SIZE 15, STERILE; BARD-PARKER BLADE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASPEN SURGICAL PRODUCTS, CALEDONIA BARD-PARKER CARBON RIB-BACK BLADES SIZE 15, STERILE; BARD-PARKER BLADE Back to Search Results
Model Number 371115-150
Device Problem Component Falling (1105)
Patient Problem Foreign Body In Patient (2687)
Event Date 02/01/2018
Event Type  Injury  
Manufacturer Narrative
The distributor indicated in report that no sample was available for evaluation, only the manufacturing lot number was provided for review.Particles from a bard-parker blade size 15 was reported to have fallen into the wound during a bunionectomy.The surgeon irrigated the wound to flush out any particles from the blade.The patient tolerated well and follow up care was not required.Based on the evaluation conducted, lot 0146444 was packed per dhr and no non-conformance related to this issue were present.Packing process has established controls to mitigate the condition, including a blade sensor that inspect 100% of packed pouches liner level prior to aluminum foil packaging.The following controls are in-place to mitigate ¿broken blade¿ condition at aspen surgical (b)(4) site: heat treatment in-process at the beginning and end of each lot are inspected for dimension integrity using a pin gauge, flatness gauge and perforation length gauge, ductility test, and hardness test.Heat treatment quality inspections at the beginning and end of each lot are inspected for dimension integrity using a pin gauge, flatness gauge and perforation length gauge, ductility test, and hardness test.Based on this information, no further action is required.
 
Event Description
Aspen surgical received a report from the distributor indicating that particles from a bard-parker blade fell in the patient during a procedure.The incident occurred at the user facility.This report was filed in our complaint handling system as (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD-PARKER CARBON RIB-BACK BLADES SIZE 15, STERILE
Type of Device
BARD-PARKER BLADE
Manufacturer (Section D)
ASPEN SURGICAL PRODUCTS, CALEDONIA
6945 southbelt dr. s.e.
caledonia MI 49316
Manufacturer (Section G)
SAME AS ABOVE
Manufacturer Contact
jordan hackert
6945 southbelt dr. s.e.
caledonia, MI 49316
6165367508
MDR Report Key7316765
MDR Text Key101606772
Report Number1836161-2018-00030
Device Sequence Number1
Product Code GES
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 02/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number371115-150
Device Lot Number0146444
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/06/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-