Model Number S7 |
Device Problems
No Display/Image (1183); Device Stops Intermittently (1599)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/09/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information not provided due to (b)(6) patient privacy regulations.No procode, common device name, unique device identification (udi) and/or 510k provided as this device is not released for distribution in the united states.Device manufacturing date is unavailable.A medtronic representative went to the site to test the equipment.It was reported that the monitor of the navigation system was replaced.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts have been received by the manufacturer for evaluation.
|
|
Event Description
|
A site representative reported that, while in a cranial resection, the surgeon monitor of the navigation system lost display functionality shortly after connecting it to the navigation system.It was noted that the cable would be re-connected to restore functionality.The surgeon completed the procedure with the use of the navigation system.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no impact on patient outcome.No additional information was provided.
|
|
Manufacturer Narrative
|
Additional information: device manufacture date provided.
|
|
Manufacturer Narrative
|
Correction: medtronic received information that the monitor of the navigation system was not replaced as the site did not provide permission to replacement the component.If information is provided in the future, a supplemental report will be issued.
|
|