Device Problem Suction Problem (2170)
Patient Problem No Patient Involvement (2645)
Event Date 02/09/2018
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The internal suction tubing was adjusted to remove kinks that were causing the suction failure.No report of patient involvement.Unique device identifier (b)(4).The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported that the suction is not working on the anesthetic machine.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 650 A1
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi
CH 
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7320158
MDR Text Key101860634
Report Number9710602-2018-00063
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial
Report Date 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberSEE BLOCK H10
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-