MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 02/14/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern. Device manufacturing date is unavailable. A medtronic representative went to site to the system. Representative reported that a monitor was replaced to resolve the issue. A system checkout was performed and the hardware, software, and instruments passed the system checkout. The system was found to be fully functional. The suspect monitor has not been received by the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that while outside of procedure, during application the main monitor lost power but the secondary monitor was still functioning. When the hdmi cable was connected to secondary monitor it was working normally. Main cart did not display no signal or green or red dot. It was reported that there was no audio from system. There was no patient present at the time of the issue.
 
Manufacturer Narrative
Additional information: device manufacturer date provided. Correction: product, unique device identification (udi) and associated fields updated to proper value.
 
Manufacturer Narrative
The monitor for the navigation system was returned to the manufacturer for evaluation. Testing found that the led indicator light would not power on when power was applied to the unit. It was noted that the state of the monitor did not change. The hardware investigation found that the reported event was related to a hardware issue. This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S8 PREMIUM SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7324645
MDR Text Key101997064
Report Number1723170-2018-01057
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 07/23/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number9735665
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/17/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-