DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED
|
Back to Search Results |
|
Catalog Number 532.010 |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Reporter¿s phone number: (b)(6).
The actual device was returned for evaluation.
Reliability engineering evaluated the device and determined that the device would not run.
It was also determined that the device failed pre-test for check the off/oscillation/on switch mode function since the device did not run and when the trigger was pressed there was no function.
Therefore, the reported condition was confirmed.
The assignable root cause was determined to be due to improper cleaning and maintenance.
If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the small battery drive stopped working.
It was not reported if there were any delays in the surgical procedure or whether a spare device was available for use.
There was patient involvement.
There were no injuries, medical intervention or prolonged hospitalization.
The exact date of this event is unknown.
All available information has been disclosed.
If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|