During processing of this complaint, attempts were made to obtain complete event, patient and device information.
Evaluation summary: visual inspection was performed on the returned device.
The reported separation was confirmed.
The reported difficulty positioning the device with the guiding catheter could not be replicated in a testing environment as it was based on operational circumstances.
A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.
Additionally, a review of the complaint history identified no other similar incidents from this lot.
The investigation determined the reported difficulties appear to be related to circumstances of the procedure.
There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|