The gas blender system 25-40-00 is not distributed in the usa, but it is similar to gas blender system 25-40-45, which is distributed in the usa (510(k) number: k052601).
This event was initially considered to be non-reportable.
However, after additional evaluation, livanova (b)(4) has decided to reclassify this event as reportable.
This report is being filed as part of a retrospective review conducted in response to this new decision.
Livanova (b)(4) manufactures the s3 gas blender system.
The incident occurred in (b)(6).
The device was sent to livanova (b)(4) for repair.
The reported failure code could not be confirmed.
Instead another error message was detected by a livanova service engineer during the investigation.
The root cause can be traced back to a defective dc-supply.
The customer decided that no repair should be performed and the device will be scrapped.
A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
|